Penumbra Announces Worldwide Voluntary Field Removal of the Original Version of 6F Neuron 070 Delivery Catheter

FOR IMMEDIATE RELEASE - September 29, 2009 - ALAMEDA, CA - (BUSINESS WIRE)-

Penumbra is initiating a voluntary Field Removal of the original version of Neuron 6F 070 Delivery Catheter. For these catheters, Penumbra has received feedback from some users that the catheter could kink or ovalize in certain anatomical situations. This kinking and/or ovalization of the catheter lumen can lead to difficulty in catheter advancement and/or delivery of other devices through the guide catheter. Continued usage of a kinked catheter could cause the catheter to break, which could result in a portion being retained in the patient, which could result in additional procedural time, patient injury or death.

Consumers who have the older version of 6F Neuron 070 Delivery Catheter which is being recalled should stop using and return the device. If your facility is in possession of units with Lot numbers F14630 or higher, you may continue to use these units. Units with Lot numbers F14630 or higher are not affected by this Field Removal Action.

The following product catalog numbers and Lot numbers are affected by this Field Removal. Neuron Delivery Catheter 070 catalog number: PND6F0701058, PND6F0701058M, PND6F070956 and PND6F070956M. Affected Lot numbers include: F13733, F13751, F13773, F13799, F13817, F13826, F13856, F13865, F13880, F13899, F13934, F13960, F13970, F13971, F13980, F13997, F14007, F14028, F14045, F14046, F14047, F14048, F14102, F14103, F14150, F14182, F14183, F14427, F14482, F14560 and F14629.

Penumbra voluntarily conducted a field removal of the devices after learning about the devices potential to kink or ovalize. FDA has been apprised of this action.

There are four injury reports to date, which may be potentially associated with this defect.

Penumbra is notifying affected hospitals through detailed Field Removal notification letters requesting that affected product be returned to Penumbra. Replacement product will be available for all affected customers. Consumers with questions may contact Penumbra directly at 1-888-272-4606 or 510-748-3200.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax

Penumbra Catheter Injury Lawsuit

If you or a loved one have have been injured, had a stoke, heart attack or any of serious injury as a result of a recalled or defective Penumbra Catheter, then call us to discuss your legal rights to monetary compensation for your injuries, stroke, heart attack or the death of a loved one due to a Penumbra catheter. The Willis Law Firm has represented thousands of clients across the United States in defective and recalled medical devices, recalled drugs and other product liability claims. Mr. Willis is a Board Certified Trial Lawyer with 25+ years of legal experience. Call us at 1-800-883-9858 or fill out our online case form.

Posted by: AT 07:11 pm   |  Permalink   |  Email
Willis Law Firm - Houston, Texas Drug Injury Lawyer and Attorney Free Lawsuit & Class Action Lawsuit Lawyer

 

LEGAL ADVERTISING

THIS WEBSITE IS NOT A MEDICAL SITE

This is an Attorney / Lawyer Advertising Website. It is not a medical information, medical dictionary, medical advice or medical help site. Talk to you own treating Doctor about any questions or concerns that you have concerning any medicine, drug, medical device or treatment plan. Do not rely on any information on this site to alter, stop or change any medication or treatment plan by your prescribing or treating Doctor. Many of the drugs and devices listed here on this site have published risk and warnings associated with their use and have not been recalled. This website is not in anyway connected with any drug company or the FDA.Check with your Doctor, Pharmacy and/or the FDA as to any concerns as to their use, recall status and safety. See http://www.fda.gov/ for the latest information.

Principal Office -Willis Law Firm 5005 Riverway Drive, Suite #160 Houston, Texas 77056

____________

.

.

.

.

.